CRP Calibrator
|
|
Cat. No.:
|
KAI-012C
|
|
Type.:
|
Calibrator
|
|
|
KIT COMPOSITION
|
|
Calibrator A
|
150 mM Sodium Chloride
|
1 x 1 mL
|
|
Calibrator B-F
|
Human CRP
|
5 x 1 mL
|
|
|

Click Image To Enlarge
|
|
|
INTENDED USE
|
The K-ASSAY ®
CRP Calibrator is intended to be used for the calibration of the
K-ASSAY ®
CRP immunoturbidimetric assay (Cat. No. KAI-026) for quantifying serum levels of C-Reactive Protein (CRP).
FOR IN VITRO DIAGNOSTIC USE.
|
|
CALIBRATOR PREPARATION
|
|
The calibrators are ready to use. They do not require reconstitution.
|
|
|
STORAGE AND HANDLING
|
|
All calibrators should be stored refrigerated (2-8°C). Return all calibrators
to 2-8°C promptly after use. Unopened calibrators can be used for one year
from the date of manufacturing as indicated on the expiration date on the package and
bottle labels. Opened bottles of calibrators can be used for 1 month if stored at 2-8°C.
|
|
CALIBRATOR VALUES
|
The following values are given as an example. The expected values for the
K-ASSAY ®
CRP Calibrator are continually being revised through ongoing quality assurance.
As a result, the expected values may change from lot to lot. Please refer to the
package insert for each lot for the exact calibrator values.
| Cal Level |
mg/dL |
mg/L |
µg/dL |
| A |
0 |
0 |
0 |
| B |
0.9 |
9 |
900 |
| C |
4.1 |
41 |
4,100 |
| D |
8.0 |
80 |
8,000 |
| E |
12.2 |
122 |
12,200 |
| F |
23.9 |
239 |
23,900 |
This calibrator has been standardized against CAP/BCR/IFCC Reference Preparation for
Proteins in Human Serum (RPPHS), lot 91/0619.
|