Apo AII / CII / CIII Calibrator
|
|
Cat. No.:
|
KAI-041C
|
|
Type.:
|
Calibrator
|
|
|
KIT COMPOSITION
|
|
Lyophilized Human Serum
|
1 x 1 mL
|
|
Diluent (saline solution)
|
2 x 2 mL
|
|
|

Click Image To Enlarge
|
|
|
INTENDED USE
|
The K-ASSAY ®
Apo AII / CII / CIII Calibrator is intended to be used for the multi-point calibration of the
K-ASSAY ®
Apo AII (Cat. No. KAI-003), Apo CII (Cat. No. KAI-005), and Apo CIII (Cat. No. KAI-006) immunoturbidimetric assays.
FOR RESEARCH USE ONLY IN THE U.S.
NOT FOR USE IN DIAGNOSTIC PROCEDURES IN THE U.S.
|
|
CALIBRATOR PREPARATION
|
|
1.
|
Add exactly 1 mL of purified water to the vial containing the lyophilized serum.
Swirl gently to dissolve.
|
|
|
2.
|
Prepare 5 calibrators (B-F) using the dissolved serum and diluent. See product
package insert for details.
|
|
|
STORAGE AND HANDLING
|
|
All calibrators should be stored refrigerated (2-8°C). Return all calibrators
to 2-8°C promptly after use. Unopened calibrators can be used for one year
from the date of manufacturing as indicated by the expiration date on the package and
bottle labels. Opened, reconstituted, and diluted calibrators can be used for 2
weeks if stored at 2-8°C.
|
|
CALIBRATOR VALUES
|
The following values are given as an example. The expected values for the
K-ASSAY ®
Apo AII / CII / CIII Calibrator are continually being revised through ongoing quality assurance.
As a result, the expected values may change from lot to lot. Please refer to the
package insert for each lot for the exact calibrator values.
Concentration of lyophilized serum, reconstitued with 1 mL of purified water (undiluted) :
|
Apo AII
|
Apo CII
|
Apo CIII
|
|
94 mg/dL
|
13.3 mg/dL
|
30.5 mg/dL
|
After reconstitution and dilution :
|
Calibrators A-F, values given in mg/dL
|
|
A |
B |
C |
D |
E |
F |
| Apo AII |
0.0 |
9.4 |
18.8 |
37.6 |
56.4 |
94.0 |
| Apo CII |
0.0 |
1.3 |
2.7 |
5.3 |
8.0 |
13.3 |
| Apo CIII |
0.0 |
3.1 |
6.1 |
12.2 |
18.3 |
30.5 |
|