Immunoassay Calibrators
 




CRP(2) Calibrator Set A
Cat. No.: KAI-061C
Type.: Calibrator


KIT COMPOSITION
Calibrator 1-5 Human CRP 5 x 2 mL
    
KAI-061C
Click Image To Enlarge       


INTENDED USE
The K-ASSAY ®   CRP(2) Calibrator Set A is intended to be used for the calibration of the K-ASSAY ®   CRP(2) immunoturbidimetric assay (Cat. No. KAI-060) for quantitating C-Reactive Protein (CRP) in serum samples.
FOR IN VITRO DIAGNOSTIC USE.

CALIBRATOR PREPARATION
The calibrators are ready to use.  They do not require reconstitution.

STORAGE AND HANDLING
All calibrators should be stored refrigerated (2-8°C).  Return all calibrators to 2-8°C promptly after use.  Unopened calibrators can be used for 18 months from the date of manufacturing as indicated on the expiration date on the package and bottle labels.  Opened bottles of calibrators can be used for 1 month if stored at 2-8°C.

CALIBRATOR VALUES
The following values are given as an example.  The expected values for the K-ASSAY ®   CRP(2) Calibrator Set A are continually being revised through ongoing quality assurance.   As a result, the expected values may change from lot to lot.  Please refer to the package insert for each lot for the exact calibrator values.

  Cal #     mg/dL     mg/L     µg/dL  
  1     0.050     0.50     50  
  2     0.150     1.50     150  
  3     0.500     5.00     500  
  4     1.000     10.00     1,000  
  5     2.000     20.00     2,000  

Note: Isotonic saline (not included) should be used for Calibrator 0. (0.00 mg/dL)

This calibrator is traceable to CAP/BCR/IFCC Reference Preparation for Proteins in Human Serum (RPPHS), lot 91/0619.