CRP(2) Calibrator Set B
|
|
Cat. No.:
|
KAI-062C
|
|
Type.:
|
Calibrator
|
|
|
KIT COMPOSITION
|
|
Calibrator 1-5
|
Human CRP
|
5 x 2 mL
|
|
|

Click Image To Enlarge
|
|
|
INTENDED USE
|
The K-ASSAY ®
CRP(2) Calibrator Set B is intended to be used for the calibration of the
K-ASSAY ®
CRP(2) immunoturbidimetric assay (Cat. No. KAI-060) for quantitating C-Reactive Protein (CRP) in serum samples.
FOR IN VITRO DIAGNOSTIC USE.
|
|
CALIBRATOR PREPARATION
|
|
The calibrators are ready to use. They do not require reconstitution.
|
|
|
STORAGE AND HANDLING
|
|
All calibrators should be stored refrigerated (2-8°C). Return all calibrators
to 2-8°C promptly after use. Unopened calibrators can be used for 18 months
from the date of manufacturing as indicated on the expiration date on the package and
bottle labels. Opened bottles of calibrators can be used for 1 month if stored at 2-8°C.
|
|
CALIBRATOR VALUES
|
The following values are given as an example. The expected values for the
K-ASSAY ®
CRP(2) Calibrator Set B are continually being revised through ongoing quality assurance.
As a result, the expected values may change from lot to lot. Please refer to the
package insert for each lot for the exact calibrator values.
| Cal # |
mg/dL |
mg/L |
µg/dL |
| 1 |
0.050 |
0.50 |
50 |
| 2 |
0.150 |
1.50 |
150 |
| 3 |
0.500 |
5.00 |
500 |
| 4 |
0.750 |
7.50 |
750 |
| 5 |
1.000 |
10.00 |
1,000 |
Note: Isotonic saline (not included) should be used for Calibrator 0. (0.00 mg/dL)
This calibrator is traceable to CAP/BCR/IFCC Reference Preparation for
Proteins in Human Serum (RPPHS), lot 91/0619.
|