Direct HDL-Cholesterol Calibrator
Direct HDL-C Calibrator
|
|
Cat. No.:
|
KHD-C01 or KHD-C02
|
|
Type.:
|
Calibrator
|
|
|
KIT COMPOSITION
|
Cat. No. KHD-C01
|
Lyophilized Human Serum
|
4 x 1 mL
|
|
( Note all vials are the same concentration. )
|
|
|
Image Not Available
|
|
Cat. No. KHD-C02
|
Lyophilized Human Serum
|
1 x 1 mL
|
|
|

Click Image To Enlarge
|
|
|
INTENDED USE
|
For the calibration of the
K-ASSAY ®
Direct HDL-Cholesterol assay.
FOR IN VITRO DIAGNOSTIC USE.
|
|
CALIBRATOR PREPARATION
|
|
1.
|
Using a Class A volumetric pipette, carefully add exactly 1 mL of distilled or deionized water into the vial of the
calibrator material.
|
|
|
2.
|
Swirl vial gently and allow to sit at room temp. for 5-10 min.
|
|
|
3.
|
Swirl contents until homogeneous. DO NOT VORTEX.
Remove the sample required for testing.
|
|
|
STORAGE AND HANDLING
|
The unreconstituted Direct HDL-Cholesterol calibrator is stable until the expiration
date stated on the vial label when stored at 2-8°C protected from light.
Once reconstituted, the calibrator material is stable for up to 5 days when stored at
2-8°C. DO NOT FREEZE the calibrator once reconstituted.
|
|
CALIBRATOR VALUES
|
The following values are given as an example. The expected values for the
K-ASSAY ®
Direct HDL-Cholesterol Calibrator are continually being revised through ongoing quality assurance.
As a result, the expected values may change from lot to lot. Please refer to the
package insert for each lot for the exact calibrator values.
|