Plasma FDP Calibrator
Cat. No.: KAI-112C
Type: Calibrator


KIT COMPOSITION
Lyophilized Human Plasma FDP 5 x 1 mL
Diluent (Tris-HCl solution) 1 x 40 mL
    
KAI-112C
Click Image To Enlarge       

Click here for an example package insert.

Click here for the Safety Data Sheet (SDS).


INTENDED USE
The K-ASSAY ®   Plasma FDP Calibrator is intended to be used for the multi-point calibration of the K-ASSAY ®   Plasma FDP assay (Cat. No. KAI-111).
FOR RESEARCH USE ONLY IN THE U.S.
NOT FOR USE IN DIAGNOSTIC PROCEDURES IN THE U.S.

CALIBRATOR PREPARATION
 1.  Add exactly 1.0 mL of diluent to a vial containing the lyophilized Factor XIII to dissolve it and allow to sit for 30 minutes.   Swirl gently to dissolve.

 2.  Prepare 6 calibrators (A-F) using the dissolved stock and diluent.  See product package insert for details.

STORAGE AND HANDLING
All stock calibrators and diluent should be stored refrigerated (2-8°C).  Return all calibrators to 2-8°C promptly after use.

Unopened calibrators can be used for 1 year from the date of manufacturing as indicated by the expiration date on the package and bottle labels.

Reconstituted calibrators can be used for 6 hours if stored at 2-8°C.

Diluted calibrators can be used for 2 hours if stored at 2-8°C.

Reconstituted stock calibrators (not diluted) can be frozen one time and stored at -20°C for up to 1 month.  Do not refreeze a second time.

Unopened diluent can be used for 2 years from the date of manufacturing as indicuated by the expiration date on the package and bottle labels.  Opened diluent can be used up to the expiration date provided that the cap is tightly closed to prevent evaporation and the solution is not contaminated.

CALIBRATOR VALUES
The following values are given as an example.  The expected values for the K-ASSAY ®   Plasma FDP Calibrator are continually being revised through ongoing quality assurance.   As a result, the expected values may change from lot to lot.  Please refer to the package insert for each lot for the exact calibrator values.

Concentration of lyophilized Plasma FDP, reconstitued with 1.0 mL of diluent :
80 µg/mL

After reconstitution and dilution :
Calibrators A-F
( µg/mL )
A B C D E F
  0.0     5.0     10.0     20.0     40.0     80.0