Cat. No. KAI-005

Apo CII

For Research Use Only in the U.S.
KIT COMPOSITION

Reagent 1 (R-1) Buffer Reagent 3 x 20 ml
Reagent 2 (R-2) Antiserum Reagent 1 x 20 ml


PERFORMANCE

Sensitivity

When a saline blank is used as a sample, the absorbance is below 0.200. When a standard having an apolipoprotein CII concentration of around 4 mg/dl is assayed, the absorbance (after subtracting the saline blank) is within 0.010-0.100.

Specificity

When control serum with a known value is assayed, the result is within 10% of the assigned value.

Assay Range

1-15 mg/dl (multi-point calibration)

Precision

When a sample that has an apolipoprotein CII concentration of around 4 mg/dl is assayed 20 times (within-run), the absorbance C.V. is below 5%.

Correlation

Y
 r
 n
 X 
 Y 
=  0.817 X + 0.642
=  0.9915
=  54
=  company A's ITA
K-ASSAY ITA APO C II


INTERFERENCE

Bilirubin
Hemoglobin
Lipemia
No interference up to 20 mg/dl
No interference up to 500 mg/dl
No interference up to 20%