KAMIYA BIOMEDICAL COMPANY
   25 Years!
Products  
 Search All Products
 
 
 Antibodies
 - A B C D E F G H I J K
 - L M N O P Q R S T U
 - V W X Y Z  Other
 Biochemicals
 - A B C D E F G H I J K
 - L M N O P Q R S T U
 - V W X Y Z  Other
 
 Assay Kits
 
 
 Research Interests 
 - Cancer
 - Cardiac
 
 Antibody Applications 
 - ELISA
 

Cat. No. KAI-009

Complement C3

For In-Vitro Diagnostic Use
KIT COMPOSITION

Reagent 1 (R-1) Buffer Reagent 4 x 20 ml
Reagent 2 (R-2) Antiserum Reagent 2 x 10 ml


PERFORMANCE

Precision

The precision for the K-ASSAY C3 assay was determined using packaged reagents, pooled human serum, and a Hitachi 717 analyzer.

Precision Assay: Within Run
Sample I Sample II Sample III
N = 20
Mean = 33.7
SD = 1.12
CV = 3.33%
N = 20
Mean = 75.6
SD = 1.68
CV = 2.22%
N = 20
Mean = 112.4
SD = 2.18
CV = 1.94%

Precision Assay: Between Runs
Sample I Sample II Sample III
N = 10
Mean = 35.4
SD = 1.29
CV = 3.63%
N = 10
Mean = 84.3
SD = 1.95
CV = 2.31%
N = 10
Mean = 114.4
SD = 1.70
CV = 1.48%

Correlation

A comparison of the K-ASSAY C3 assay and an Incstar C3 Test Kit was performed using a Hitachi 717. The test results provided the following data:

Y
 r
 n
 X 
 Y 
=  0.842X + 4.727
=  0.925
=  60
=  Incstar C3 Test Kit
K-ASSAY C3 assay
 
X min  =   78 Y min  =   74
   max =  188    max =  171
   mean  =  135    mean =  118

Assay Range

30-350 mg/dl (multi-point calibration)


INTERFERENCE

Bilirubin F and C
Hemoglobin
Lipemia
No interference up to 20 mg/dl
No interference up to 500 mg/dl
No interference up to 5%

⇐   BACK
12779 Gateway Drive, Seattle, WA 98168, USA  Tel: (206) 575-8068  Fax: (206) 575-8094  E-mail: webmaster@kamiyabiomedical.com