Cat. No. KAI-010

Complement C4

For In-Vitro Diagnostic Use
KIT COMPOSITION

Reagent 1 (R-1) Buffer Reagent 4 x 20 ml
Reagent 2 (R-2) Antiserum Reagent 2 x 10 ml


PERFORMANCE

Precision

The precision for the K-ASSAY C4 assay was determined using packaged reagents, pooled human serum, and a Hitachi 717 analyzer.

Precision Assay: Within Run
Sample I Sample II Sample III
N = 20
Mean = 5.7
SD = 0.29
CV = 5.17%
N = 20
Mean = 36.3
SD = 0.76
CV = 2.1%
N = 20
Mean = 58.6
SD = 1.98
CV = 3.38%

Precision Assay: Between Runs
Sample I Sample II Sample III
N = 10
Mean = 6.51
SD = 0.24
CV = 3.72%
N = 10
Mean = 30.6
SD = 1.18
CV = 3.87%
N = 10
Mean = 53.2
SD = 1.12
CV = 2.1%

Correlation

A comparison of the K-ASSAY C4 assay and an Incstar C4 Test Kit was performed using a Hitachi 717. The test results provided the following data:

Y
 r
 n
 X 
 Y 
=  0.739X + 2.46
=  0.988
=  60
=  Incstar C4 Test Kit
K-ASSAY C4 assay
 
X min  =   11 Y min  =    9
   max =   56    max =   45
   mean  =   31    mean =   25

Assay Range

8-80 mg/dl (multi-point calibration)


INTERFERENCE

Bilirubin F and C
Hemoglobin
Lipemia
No interference up to 20 mg/dl
No interference up to 500 mg/dl
No interference up to 5%

⇐   BACK