Cat. No. KAI-022

Haptoglobin

For In-Vitro Diagnostic Use
KIT COMPOSITION

Reagent 1 (R-1) Buffer Reagent 3 x 18 ml
Reagent 2 (R-2) Antiserum Reagent 2 x 7 ml


PERFORMANCE

Precision

The precision for the K-ASSAY HP assay was determined using packaged reagents, pooled human serum, and a Hitachi 717 analyzer.

Precision Assay: Within Run
Sample I Sample II
N = 20
Mean = 68.15
SD = 0.88
CV = 1.3%
N = 20
Mean = 366
SD = 5.5
CV = 1.5%

Precision Assay: Between Runs
Sample I Sample II
N = 8
Mean = 85.9
SD = 0.8
CV = 0.972%
N = 8
Mean = 225.6
SD = 1.4
CV = 0.624%

Correlation

A comparison the the K-ASSAY HP assay and an Incstar Haptoglobin Test Kit was performed using a Hitachi 717. The test results provided the following data:

Y
 r
 n
 X 
 Y 
=  0.937X + -9.25
=  0.993
=  60
=  Incstar Haptoglobin Test Kit
K-ASSAY HP assay
 
X min  =   34 Y min  =    4
   max =   26    max =  267
   mean  =  109    mean =   93

INTERFERENCE

Bilirubin F and C
Hemoglobin
Lipemia
No interference up to 20 mg/dl
No interference up to 500 mg/dl
No interference up to 5%