Cat. No. KAI-023

Transferrin

For In-Vitro Diagnostic Use
KIT COMPOSITION

Reagent 1 (R-1) Buffer Reagent 3 x 20 ml
Reagent 2 (R-2) Antiserum Reagent 2 x 10 ml


PERFORMANCE

Precision

The precision for the K-ASSAY Transferrin assay was determined using packaged reagents, pooled human serum, and a Hitachi 717 automatic analyzer.

Precision Assay: Within Run
Sample I Sample II Sample III
N = 20
Mean = 70.7
SD = 0.92
CV = 1.3%
N = 20
Mean = 300.75
SD = 5.02
CV = 1.67%
N = 20
Mean = 526.4
SD = 5.77
CV = 1.097%

Precision Assay: Between Runs
Sample I Sample II
N = 8
Mean = 175.38
SD = 1.6
CV = 0.91%
N = 8
Mean = 340.9
SD = 1.46
CV = 0.4%

Correlation

A comparison of the K-ASSAY Transferrin assay to an Incstar Transferrin Test Kit was performed using a Hitachi 717. The test results provided the following data:

 

Y
 r
 n
 X 
 Y 
=  1.064 X + -50.124
=  0.982
=  60
=  Incstar Trasferrin Test Kit
K-ASSAY Transferrin assay
 
X min  =   148 Y min  =   110
   max =  440    max =  426
   mean  =   246    mean =   212
 

INTERFERENCE

Ascorbic Acid
Bilirubin F and C
Hemoglobin
Lipemia
No interference up to 50 mg/dl
No interference up to 20 mg/dl
No interference up to 500 mg/dl
No interference up to 5%

⇐   BACK