Cat. No. KAI-031

RF

For In-Vitro Diagnostic Use
KIT COMPOSITION

Reagent 1 (R-1) Buffer Reagent 4 x 20 ml
Reagent 2 (R-2) RF Reagent 2 x 8 ml


PERFORMANCE

Precision

The precision for the K-ASSAY Rheumatoid Factor was determined using packaged reagents, pooled human serum, and a Hitachi 717 automatic analyzer.

Precision Assay: Within Run
Sample I Sample II
N = 20
Mean = 3.3
SD = 0.86
CV = 26.197%
N = 20
Mean = 110.15
SD = 1.5
CV = 1.359%

Precision Assay: Between Runs
Sample I Sample II Sample III
N = 10
Mean = 7.9
SD = 2.13
CV = 26.98%
N = 10
Mean = 33
SD = 1.33
CV = 4.04%
N = 10
Mean = 106.3
SD = 2.06
CV = 1.94%

Assay Range

5-320 IU/ml (multi-point calibration)


Correlation

Y
 r
 n
 X 
 Y 
=  0.876x - 20.268
=  0.984
=  43
=  company A's ITA
K-ASSAY Rheumatoid Factor
 

INTERFERENCE

Bilirubin F and C
Hemoglobin
Lipemia
No interference up to 20 mg/dl
No interference up to 500 mg/dl
No interference up to 5%