Cat. No. KAI-035

Fibrinogen

For In-Vitro Diagnostic Use
KIT COMPOSITION

Reagent 1 (R-1) Buffer Reagent 2 x 20 ml
Reagent 2 (R-2) Antiserum Reagent 1 x 8 ml
Fibrinogen Calibrator (lyophilized) 1 x 1 ml


PERFORMANCE

Sensitivity

When a saline blank is used as a sample, the absorbance is below 0.050. When a standard having a fibrinogen concentration of around 248 mg/dl is assayed, the absorbance (after subtracting the saline blank) is within 0.050-0.150.

Specificity

When control serum with a known value is assayed, the result is within 10% of the assigned value.

Assay Range

100-900 mg/dl (multi-point calibration) or
81-729 mg/dl (using NIBSC international standard)

Precision

When a sample containing 200 mg/dl fibrinogen is assayed 20 times (within-run), the absorbance C.V. is below 5%.

Within Run Sample A Sample B Sample C
N
MAX
MIN
AVE
CV%
20
152
136
142.4
2.96
20
456
429
444.7
1.48
20
763
721
739.6
1.43

Between Day Sample A Sample B Sample C
N
MAX
MIN
AVE
CV%
10
170
165
167.4
0.94
10
329
316
323.6
1.06
10
571
544
554.1
1.50

Correlation

A comparison ofthe K-ASSAY Fibrinogen test and an Incstar Fibrinogen test kit were performed on a Hitachi 717. The test results provided the following data:

Y
 r
 n
 X 
 Y 
=  0.967x + 33.91
=  0.995
=  50
=  Incstar ITA
K-ASSAY Fibrinogen

Additional correlation sutdies on assays using the Clauss method (clotting method) are available.