KAMIYA BIOMEDICAL COMPANY
   25 Years!
Products  
 Search All Products
 
 
 Antibodies
 - A B C D E F G H I J K
 - L M N O P Q R S T U
 - V W X Y Z  Other
 Biochemicals
 - A B C D E F G H I J K
 - L M N O P Q R S T U
 - V W X Y Z  Other
 
 Assay Kits
 
 
 Research Interests 
 - Cancer
 - Cardiac
 
 Antibody Applications 
 - ELISA
 

Cat. No. KAI-036

Plasminogen

For In-Vitro Diagnostic Use
KIT COMPOSITION

Reagent 1 (R-1) Buffer Reagent 4 x 20 ml
Reagent 2 (R-2) Antiserum Reagent 2 x 10 ml
Plasminogen Calibrator (lyophilized) 1 x 1 ml


PERFORMANCE

Precision

The precision for the K-ASSAY plasminogen assay was determined using packaged reagents, pooled human plasma, and a Hitachi 717 analyzer.

Precision Assay: Within Run
Sample I Sample II
N = 20
Mean = 5.33
SD = 0.086
CV = 1.62%
N = 20
Mean = 11.4
SD = 0.122
CV = 1.07%

Precision Assay: Between Runs
Sample I Sample II
N = 10
Mean = 5.31
SD = 0.074
CV = 1.39%
N = 10
Mean = 11.5
SD = 0.099
CV = 0.86%

Correlation

A comparison of the K-ASSAY Plasminogen assay and a Binding Site Plasminogen RID Kit was performed using a Hitachi 717. The test results provided the following data:

 

Y
 r
 n
 X 
 Y 
=  0.995 X + 0.309
=  0.978
=  39
=  Binding Site Plasminogen RID Kit
K-ASSAY Plasminogen assay
 
X min  =    3.9 Y min  =   4.0
   max =   17.4    max =   17.6
   mean  =   10.9    mean =   11.2



INTERFERENCE

Bilirubin F and C
Hemoglobin
Lipemia
No interference up to 20 mg/dl
No interference up to 500 mg/dl
No interference up to 5%

⇐   BACK