Cat. No. KAI-053

Prealbumin

For In-Vitro Diagnostic Use
KIT COMPOSITION

Reagent 1 (R-1) Buffer Reagent 4 x 20 ml
Reagent 2 (R-2) Antiserum Reagent 2 x 8 ml


PERFORMANCE

Precision

The precision for the K-ASSAY Prealbumin assay was determined using packaged reagents, pooled human serum, and a Hitachi 717 analyzer.

Precision Assay: Within Run

Sample I Sample II Sample III
N = 20
Mean = 11.2
SD = 0.145
CV = 1.29%
N = 20
Mean = 24.9
SD = 0.289
CV = 1.16%
N = 20
Mean = 57.0
SD = 0.767
CV = 1.35%

Correlation

A comparison of the K-ASSAY Prealbumin assay and a similar Prealbumin assay was performed on a Hitachi 717. The test results provided the following data.

Y
 r
 n
 X 
 Y 
=  0.9164x + -0.332
=  0.9924
=  50
=  Similar Prealbumin assay
K-ASSAY Prealbumin assay
 
X min  =    5.7 Y min  =    4.2
   max =  43.2    max =   38.4
   mean  =   22.89    mean =   20.64

INTERFERENCE

Bilirubin F and C
Hemoglobin
Lipemia
No interference up to 22 mg/dl
No interference up to 470 mg/dl
No interference up to 5%