Cat. No. KAI-078

ASO (WHO)

For In-Vitro Diagnostic Use
KIT COMPOSITION

Reagent 1 (R-1) Buffer Reagent 4 x 20 ml
Reagent 2 (R-2) Latex Suspension 4 x 20 ml


PERFORMANCE

Sensitivity

When a standard having an anti-streptolysin O concentration of around 434 U/ml is assayed, the absorbance changes will be approximately 0.1-0.5..

Specificity

When control serum with a known value is assayed, the result is within 10% of the assigned value.

Precision

When a sample is assayed 20 times, the CV is less than 5%.

Assay Range

20-1,000 IU/ml

Correlation

y = 1.03x - 0.87
r = 0.994
x = company A's ASO assay
y = K-ASSAY Latex Anti-Streptolysin O